Early access · courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Library / MRI / SIGNA (1.5T family)
GE HealthCare

SIGNA (1.5T family)

The GE HealthCare SIGNA 1.5T family is a line of superconducting whole-body MRI scanners used for routine clinical imaging, including neuro, body and diffusion-weighted studies.

37 field cases from service engineers' discussions (anonymised); solved: 7, probable: 30under engineering review

02

Faults and how to find them

Diffusion (DWI) sequences fail or stop with gradient/axis errors, or show lower prescan amplitude than a healthy system. A scan may restart after an error but give poor images. Other sequences may still work.field case
Likely causes
  • Faulty gradient control board
  • Oxidised contacts in the gradient controller board connectors
  • Wear in the gradient amplifier or its power module (in one case, suspected supercapacitor wear in the lower module of the gradient cabinet)
  • Heavy load on the Z axis during diffusion, which shows up weak gradient hardware first
Checks, in order
  1. Rule out the coil by testing it on another scanner
  2. Compare prescan amplitude with a known-good system
  3. Run a different gradient-intensive protocol to see whether the fault is gradient-related
  4. Write down the exact error text (which axis, or which diffusion direction)
  5. Swap the suspect control board between the three gradient axes. If the error moves with the board, the board is faulty
  6. Swap gradient axes (for example Z with another axis) to see whether the fault follows the amplifier channel
  7. Inspect the connectors on the gradient controller boards for oxidation
Fix

In one case, replacing the gradient control board fixed the fault. This was confirmed by the error moving with the board when it was swapped between axes. In another case, reseating or swapping the gradient control boards brought diffusion back, probably by clearing oxidised contacts, but the fix may only be temporary. Where gradient amplifier or power module wear is suspected, diagnose by swapping axes and then repair or replace the affected module.

Scanning pauses or the system shuts down from heat: a scan stops for several minutes during demanding sequences (for example contrast perfusion) and then continues, or the system trips in hot weather and won't restart for about an hour, or the equipment gets noticeably hot.field case
Likely causes
  • Gradient amplifier over-temperature (one case logged OverTemperature on the Y gradient), probably because the cooling circuit was contaminated
  • Dirty chiller (lint and dust), which loses capacity in hot weather
  • Litron pump: the nylon shear bushing is sheared, so the pump shows a normal temperature but the transmit coil is not cooled
  • Room air conditioning too small for the scanner's actual heat output
Checks, in order
  1. Search the logs for OverTemperature events and note which gradient axis
  2. Check the front-panel temperature of the Litron pump and confirm the transmit coil is actually being cooled
  3. Check whether the pump's nylon shear bushing is sheared
  4. Check what coolant is in the gradient amplifier circuit (factory fluid or topped up with distilled water)
  5. Inspect and clean the chiller
  6. Use 'select viewing level' so that all error levels are shown
  7. Check that room cooling capacity matches the scanner's heat output into the equipment room and the RF room
Fix

Probable: flush the gradient cooling circuit (planned for the next scheduled maintenance), clean the chiller and do full cooling-system maintenance. If the Litron pump bushing is sheared, the advice was to replace the pump. Opening the cabinet door was used only as a temporary workaround.

Coils are not recognised or drop out: scanning fails on all coils, or during an exam the system drops the head coil and then the body coil and asks for them to be reconnected.field case
Likely causes
  • Oxidised contact in the coil connector or port on the table carriage (in one case the contact had been sitting in liquid)
  • A pin in the connector with no continuity
  • Break in the table cable harness
  • Broken contact tabs on the coil itself
Checks, in order
  1. Swap in another coil to rule out the coil
  2. Inspect the carriage coil port (port A) for oxidation and signs of liquid
  3. Check continuity on every pin of the connector
  4. Check the table cable harness for breaks
Fix

Probable: repair or replace the oxidised or open connector contact, or repair the break in the table harness. Broken coil contact tabs have been re-soldered and covered with heat-shrink, but because the coil is safety-critical this is a reviewer item (see flags).

Occasional single stripes (spikes) on images; the phantom spike-check QA fails only now and then (about 1 run in 10, usually with one spike).field case
Likely causes
  • Interference on the mains supply (in one case it started after a regional substation switchover)
  • Damaged filter on the RF plate (a coil or capacitor had broken loose inside)
  • Unshielded lighting wiring in the scan room
Checks, in order
  1. Repeat the phantom spike check
  2. Check scan room humidity, ventilation and temperature
  3. Move a small flex coil around the room in expert mode to find where the noise comes from
  4. Inspect the RF plate filters
Fix

Solved by replacing the scan room lighting wiring with shielded wiring and repairing or replacing the damaged RF plate filter. When working on the filter, hold the lower nut so the filter body does not turn.

After a reconstructor (ICN) or ICN motherboard is replaced, ICN configuration fails, installation does not pass validation, or the system does not see the ICN.field case
Likely causes
  • Configuration or installation state on the replacement ICN that does not match the system
  • ICN part number or software version that does not match the system software
Checks, in order
  1. Confirm the new ICN powers up and its iDRAC responds
  2. Check the ICN part number and that it matches the software version
  3. Connect a VGA monitor to the ICN and watch the output during boot or installation to see where it stops
  4. Reinstall the software (VRE)
Fix

In one case, a full factory reset from the reconstructor's Lifecycle Controller let the ICN configuration complete. For the motherboard-swap case the outcome was not reported. Watching the VGA output during installation was the recommended way to diagnose it.

Patient table problems: erratic table movement, or the drive keeps pushing against the cradle for about 30 s after engaging it and then stops, while later movement is normal.field case
Likely causes
  • Table power supply failure (reported as the most common failure)
  • Belt tension out of adjustment
  • Clutch out of adjustment
  • Actuator fault versus control board fault (needs to be isolated)
Checks, in order
  1. Check that the control board receives 24 V from the external supply
  2. Load-test the table power supply (about 42–48 V output, located top right in the cabinet) with an electronic load
  3. Isolate the actuator by powering it directly with 24 V from a separate source
  4. Check position readings, belt tension and clutch adjustment (in normal operation the clutch should slip for about 2–3 s)
Fix

Probable: replace the table power supply if it fails the load test. For drive-engagement overrun, adjust the belt tension and the clutch.

The system will not boot or a board does not communicate: APS does not answer ping and its LED stays solid, the AGP board in the MGD shows a 'failed' LED and does not answer ping, or a board-related fault appears.field case
Likely causes
  • Flat battery (NVRAM battery module) on a battery-backed board
  • Faulty network switch between the boards and the host
Checks, in order
  1. Ping the board and watch its status LED
  2. Use the COM port to see whether the board keeps sending requests to the host
  3. Before pulling any battery, check all connectors, including the backplane connectors
  4. Note that a software reload alone did not fix the APS case
Fix

In one case, replacing a flat board battery fixed the fault. For the AGP board, the likely fix is to replace the NVRAM battery module and then reprogram the board BOOT. Removing the battery means the board has to be reflashed. For APS not answering ping, the likely cause was the network switch, and replacing it was recommended.

Status LEDs light on the DRIVER MODULE LITE and the system no longer passes TPS reset after a reboot.field case
Likely causes
  • Failed PIN diodes in the body coil (a colleague's suggestion, not confirmed)
Checks, in order
  1. Review the system logs before replacing anything
Fix

Not confirmed. Damaged body coil PIN diodes were suggested as the likely cause; check the logs first.

Poor diffusion image quality (low SNR) with the body coil, while head and cardiac coils give acceptable results.field case
Likely causes
  • Protocol settings (b-value, NEX) that don't suit the body coil's SNR
Checks, in order
  1. Collect example images
  2. Review the protocol with the applications/clinical staff
Fix

Probable: a protocol issue rather than a hardware fault. Lowering the b-value or increasing NEX was recommended to improve SNR.

05

Field notes

For gradient faults, swapping boards or axes is the quickest isolation method: if the error moves with the board, the board is at fault.

Many DWI complaints come from gradient hardware or connector oxidation, because diffusion loads the gradients (especially Z) hardest. Rule out the coil on another system first.

Heat-related pauses and shutdowns that get worse in hot weather usually mean cooling trouble (dirty chiller, contaminated gradient coolant, failed pump). Check the logs for OverTemperature before replacing parts.

Before pulling a board's NVRAM battery, check all connectors. Removing the battery means reprogramming the board.

After replacing a reconstructor, watch its console output on a directly connected monitor during installation. It shows where configuration stops.

Helium level, cold head and cryostat issues (questionable level readings, a failed cold head, falling helium) came up in the cases, but any work affecting the magnet, cryogens or helium must be done with the manufacturer.

After any repair, run the manufacturer's full checkout and performance verification before the device goes back to patients.