Early access · courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Library / MRI / Achieva
Philips

Achieva

The Philips Achieva is a superconducting whole-body MRI scanner, built in 1.5T and 3.0T versions, used for routine and advanced clinical diagnostic imaging.

27 field cases from service engineers' discussions (anonymised); solved: 7, probable: 20under engineering review

02

Faults and how to find them

Scans stop at random, from one to a few times a day, or an optical-link error appears. The CDAS dump shows communication errors between CDAS and TXR, between PFEI and CfInt, or on the fiber, yet the PFEI and CDAS tests pass. In one case the error came and went when someone stood near the CDAS cabinet or opened its front door.field case
Likely causes
  • A faulty optical transmitter (OSDH50M2) or receiver (OEDH50M2) module on the CfInt
  • Ageing or sharply bent fiber-optic cable, especially after the system has been moved
  • A faulty transceiver on the TXR side
  • Outside light getting into the optical connections
Checks, in order
  1. Read the CDAS dump at the time of the error and note which link is reported (cdas-txr, pfei-cfint, fiber)
  2. Run the PFEI and CDAS tests. Passing tests do not rule out an intermittent optical fault.
  3. Check the fiber routing for tight bends. In a dark room, check whether the fiber lights up properly.
  4. Check the transceivers in the TXR
  5. See whether room or cabinet lighting, or opening the cabinet door, changes the behaviour
  6. Check the filter-box module and its fan
Fix

In the solved case, replacing the faulty optical module on the CfInt fixed the fault. In another case a new CfInt board fixed it for only one exam. Replacement boards and optics were then ordered, so if a new CfInt does not cure it, the optical modules and fiber are the likely cause. Run the manufacturer's checkout after any replacement.

The patient table will not move horizontally into the magnet, even in manual mode, while up/down still works. Or the table hangs and only comes back after several reboots and emergency-stop resets. In one case the /PATSUP/PLC log showed 'PLC Error 2.6 servo-ok bit not active within 1 second': the servo amplifier did not report 'servo OK' while 'Servo enable' was active.field case
Likely causes
  • Broken wires in the table cable where it bends. In the solved case, 2 of 22 wires were broken.
  • A damaged or cracked motor cable to the table drive
  • Limit switches on the bridge sticking
  • Oxidised auxiliary contacts on the contactor near the solid-state relays, at the bottom of the table
  • A broken cable running to the top of the table, to the encoder and switches
  • A faulty servo amplifier
Checks, in order
  1. Do a full reboot and check the fuses under the table
  2. Try manual mode, and check whether vertical movement still works
  3. Check the logs for PLC/servo errors
  4. Check the continuity of each wire in the table cable, especially at bend points
  5. Check the motor cable and the wiring to the limit switches inside the table
  6. Inspect the limit switches on the bridge for sticking
  7. Take apart and clean the auxiliary contactor contacts if they are oxidised
Fix

Solved case: the broken wires in the table cable were repaired. Reterminating the cable so the bend point moves is recommended. For the servo-OK error, a damaged motor cable is the likely cause, so check it and replace it if needed. For the intermittently hanging table no confirmed fix was recorded; check the switches, contacts and cable listed above. Run the manufacturer's checkout of table movement afterwards.

The eddy current test fails, with errors moving between XZ, XY or all three axes. Re-seating the cables in the gradient cabinet and restoring an old backup did not help.field case
Likely causes
  • The fill hatch on the magnet module not seating properly
Checks, in order
  1. Re-seat the interface cables, coax cables and ribbon cables in the gradient cabinet, and the axis connections
  2. Inspect the hatch on the magnet module for a proper fit
  3. Next steps that were planned: check the capacitors in the cabinet and the connections from the cabinet to the coil, and run a spike test
Fix

Solved: the eddy current test passed once the hatch was made to seat properly. Leave anything on the magnet's cryogenic side that is more than closing a cover to the manufacturer.

An intermittent error a few times a day, with the logs pointing to temperature in the gradient amplifier cooling. In a separate case, inter-layer breakdowns ('small explosions') were found in a PDM module of the gradient amplifier.field case
Likely causes
  • Problems in the gradient amplifier cooling circuit: low flow in the secondary circuits or low coolant level in the tanks
  • Thermal breakdown of the PDM module after the fan facing it failed or the air filter clogged (likely)
Checks, in order
  1. Review the detailed logs at the moment of the error
  2. Check the cooling cabinet and heat exchanger in the gradient amplifier cooling circuit
  3. Watch the amplifiers when the error occurs
  4. Check flow in the secondary circuits and the coolant level in the tanks
  5. Check that the fan facing the PDM module works and that the filter is not clogged
Fix

No confirmed fix was recorded. Overheating from poor cooling (flow, coolant, fan or filter) is the likely cause, so fix the cooling first, then replace any damaged module and run the manufacturer's checkout.

The gradient power supply (271) does not start after the system has been switched offfield case
Likely causes
  • A shorted diode in the MCD 162-12io1 module. This was likely caused by overcurrent during input voltage dips, when the thyristor is open on the rising edge and the storage capacitor has been drained by the inverter.
Checks, in order
  1. Check the MCD 162-12io1 module
  2. Review how the soft-start circuit behaves
  3. Assess the mains supply for voltage dips
Fix

No confirmed fix was recorded. The likely cause is a failed diode in the MCD 162-12io1, so replace the module and look into input voltage dips as the root cause. Run the manufacturer's checkout afterwards.

The error 'Cannot determine optimum RF power' appears when scanning with the 8-channel head coil. A rerun may work for one patient and then the error comes back with the next.field case
Likely causes
  • A broken power wire to the coil preamplifier. Preamplifier power is carried over the coax cable.
Checks, in order
  1. Rule out metal on or in the patient
  2. Run the PFEI and CDAS tests. They passed in this case, so they do not rule out a coil fault.
  3. Check the continuity of the preamplifier power path through the coil cable
Fix

The likely cause is a broken preamplifier power wire in the head coil. Repair or replace it, then verify with the coil test.

Tunnel lighting is faulty: lamps burnt out, a lamp holder missing, or LED lighting on the same line not working while the fuse under the table cover is fine. In one system, a ripple artifact over the whole image also appeared intermittently on DWI and TIR head scans (about every 3rd–4th head exam). It did not appear on body or spine exams.field case
Likely causes
  • A failed capacitor in the power supply at the bottom of the table (the outermost one), for the lighting failure (likely)
  • Interference from a faulty tunnel lamp whose broken filament made intermittent contact (suspected source of the ripple)
  • Poor contact of the PCI boards in the reconstructor
  • Poor coil connector contacts, for example after improper cleaning
Checks, in order
  1. Check the fuse under the table cover
  2. Check the tunnel lamps and their holders
  3. Check the outermost capacitor in the table power supply
  4. For artifacts: compare against the built-in Q-body coil and other anatomies to isolate the head coil
  5. Clean and degrease the coil connector pins
  6. Re-seat the PCI boards in the reconstructor
Fix

No fix here is confirmed. For dead lighting, replacing the outermost power supply capacitor at the bottom of the table is the likely fix. In the artifact case, replacing the tunnel lamps and re-seating the reconstructor PCI boards cleared the artifacts for 4 days of observation. The root cause was not confirmed, so keep monitoring and keep the head coil under suspicion.

Tests fail after the MRX board is replacedfield case
Likely causes
  • Wrong jumper settings on X20–X23. These jumpers set the board's identity (first or second board) in an 8/16-channel system.
Checks, in order
  1. Compare the X20–X23 jumper settings on the new board with the old board
Fix

Setting the jumpers to match the old board likely fixes it. In the field case the jumper was moved from X20 to X21. Rerun the tests afterwards.

The reconstructor stops working, for example overnight after a hard disk failure. Or exams take a very long time to send to PACS (up to about 15 minutes), with progress stuck at 100%.field case
Likely causes
  • A failed SAS hard disk in the reconstructor
  • A failed RAID array, with the system running on its backup array (suspected for slow PACS transfers)
  • Poor communication between the reconstructor and the host
Checks, in order
  1. Confirm the disk has failed and replace it
  2. Reboot after replacing it and check whether network boot is enabled
  3. For slow PACS transfers: check the RAID array's status and the reconstructor-to-host link
Fix

For the disk failure: replace the disk. After a full failure the reconstructor software is said to reinstall itself from the host; after a partial failure network boot may be needed (probable). For slow PACS transfers no fix was recorded; check the RAID and the reconstructor-to-host link first.

05

Field notes

Passing PFEI and CDAS tests do not rule out an intermittent optical or coil fault. Check the CDAS dump and logs at the moment of failure.

Moving the system and ageing fiber are common triggers for CDAS optical-link errors. Inspect fiber bends and check for light leaking into the optics before swapping boards.

Table cables tend to break where they repeatedly bend. After a repair, reterminate the cable so the bend point moves.

When replacing a board such as the MRX, copy the jumper settings from the old board before running tests.

A red background with 'mrl out of spec' on a PIQT report may just mean the 'out-of-tolerance values only' report type was chosen. Check the General Report before assuming a fault.

No sound from a phone plugged into the audio input: the likely fixes are pressing the button on the audio module and using a 3-contact 3.5 mm jack adapter.

After any repair, run the manufacturer's full checkout and performance verification before the device goes back to patients.