Philips

Brilliance CT

Philips Brilliance is a family of multi-slice computed tomography (CT) scanners used for routine and contrast-enhanced diagnostic imaging in hospital radiology departments.

39 field cases from service engineers' discussions (anonymised); solved: 13, probable: 26under engineering review

02

Faults and how to find them

System will not start: the key is requested or ignored, a relay clicks but the main contactor does not pull in, and the couch may not move from its buttons (it still moves by hand). The logs show nothing significant. In one case this began right after preventive maintenance.field case
Likely causes
  • Faulty slipspring e-stop assembly
  • Poor contact in the E-Stop ring/chain
  • Blown fuses or wrong switch positions under the left-hand cover
Checks, in order
  1. Walk the whole e-stop chain, including the buttons under the front and right covers and the couch stopper limit switch
  2. Check the fuses on the anode and cathode modules
  3. Check the fuses and switch positions under the left cover
  4. Check gantry tilt and confirm all CPM boards report ready
  5. Restart from the blue button and from the transformer, then review the logs
Fix

Solved in one case by repairing/replacing the faulty slipspring e-stop. In a similar post-maintenance case the likely cause was a lost contact in the E-Stop ring, with the fuses and switches under the left cover as further suspects.

Collimator errors: an A-plane collimator error stops the system from starting (the gantry window appears late at start-up), or collimator errors appear during air calibration and some exams.field case
Likely causes
  • DMC collimator A-plane parameters lost or corrupted
  • Collimator filter cracked and detached from its mount (possibly damaged during transport or installation)
  • Collimator mechanism sticking
Checks, in order
  1. Poll the circuit over CAN and check the e-stop circuit
  2. Visually inspect the collimator filter and its mount
  3. Review the density readings during calibrations
  4. Make a fresh A-plane backup before any collimator work
Fix

Solved by loading the A-plane parameters from the calibration file into the DMC and restarting. A cracked filter was likely fixed by replacing the collimator/filter after taking a fresh A-plane backup. Engineers disagreed on whether lubricating a sticking mechanism helps: the effect is likely temporary, the Brilliance collimator is considered maintenance-free, and a full calibration is needed after any disassembly.

Gantry intermittently fails to start, with RMC NO ZERO GATE logged during rotor initialisation. The index LED on the ghost board does not blink when the flag passes the optocoupler.field case
Likely causes
  • Zero gate flag misaligned, so it does not coincide with the resolver northmark pulse
  • Bent metal plate damaging or blocking the position-sensor optocoupler
Checks, in order
  1. Trace the signal in the logs and bug report
  2. Ring out the wiring from the optocoupler to the DMC
  3. Inspect the metal flag/plate for bending
  4. Swapping the optocoupler board and the DMC did not help in the solved case, so check the flag alignment first
Fix

Solved by re-adjusting the zero gate flag position so it lines up with the resolver northmark. For a similar position-sensor error, the likely fix is to check or replace the optocoupler and correct the bent plate.

After replacing the RHOST, tube warm-up and exams do not run: the warm-up countdown runs without the tube switching on, or initialisation never ends. The logs show RetZPos 0 and repeated RetZPos errors, and later the rotor spins endlessly without X-ray. Manual exposure (stationary and rotating) still works.field case
Likely causes
  • Faulty spindle block (rotor drive)
Checks, in order
  1. Reflash the RHost and restore the host backup
  2. Inspect the slip-ring contacts and brushes
  3. Check the RHost seven-segment display
  4. If the spindle block is replaced, the spindle block commissioning in Service Tools confirms the motor turns the right way
Fix

Solved by replacing the spindle block. Reflashing and restoring the backup did not help.

Scans abort, mostly on long contrast exams, with S_STARTER_FAULT_UNDER_CURRENT, S_ROTOR_STOP_ANODE_NO_MORE_SPINNING and S_GEN_MGR_SYNC_OFF_OUT_OF_SEQUENCE. A separate case showed an input overvoltage message.field case
Likely causes
  • Low mains voltage at the gantry that sags further during high-current, high-kV scans
  • Mains supply voltage out of tolerance (too high)
Checks, in order
  1. Measure the supply voltage at the gantry at rest and during a high-load scan
  2. Measure the incoming supply before the UPS and confirm your meter is reliable (one case questioned the meter battery)
  3. Clean the slip ring and brushes
Fix

Solved by tracing the problem to low mains voltage (about 440 V at the gantry, sagging to about 410 V during scans). Replacing the HSS and the generator rectifier had not helped. In the overvoltage case the system was switched off to avoid damage until the supply was corrected.

Reconstructor problems: after a system disk replacement a 'not enough resources' error appears and the full scan fails with a missing-table error (the surview still works); the RAID HDDs drop out for low speed; or the reconstructor server's motherboard does not reach BIOS.field case
Likely causes
  • Software add-on not reinstalled after the system disk was replaced
  • Worn HDDs in the RAID, possibly made worse by the power supply
  • Motherboard degradation (suspected bad solder joints)
Checks, in order
  1. Review the reconstructor errors in the logs
  2. Check the drivers and the ACQ board firmware
  3. Check the RAID disk health and the reconstructor power supply
  4. Check whether the motherboard responds to warming (this suggests solder faults)
Fix

Missing table: solved by installing the add-on. RAID: solved by replacing the HDDs with SSDs, which ran reliably for years; one suggestion was SSDs for the OS and programs, with the swap file and data on an HDD RAID. Motherboard: it started after warming, but this is a temporary sign only. Replacing the motherboard or server is the likely fix.

Image quality problems: Constancy Tests fail even though all calibrations pass; local high-density artefacts appear while bad detector shows no bad channels; groups of detector cells drop out after power-offs; or a one-off artefact appears that is gone on the repeat scan.field case
Likely causes
  • Phantom calibration out of date
  • Detector temperature not stable (detectors cooled by full shutdowns)
  • Tube kV deviation or acculon density setting
  • Transient Vicor power supply dip (for one-off artefacts)
Checks, in order
  1. Review the bad detector graphs and the date of the last phantom calibration
  2. Check the actual tube kV over CAN
  3. Measure the output with a dosimeter
  4. Check the acculon density setting and let the temperature stabilise before retesting
  5. For a one-off artefact, save the logs and take a BAG_REPORT straight away
Fix

These are likely fixes only: run a phantom calibration; adjust the acculon density, let the temperature stabilise and repeat the failed tests; for detector cell dropouts, use System Off on the console instead of a full power-off so the detectors stay warm; for a one-off artefact, a gantry reboot usually restores normal operation.

Couch faults: the couch board shows 00, then 22 after reset, and the couch does not work normally with no log errors; or errors during scanning stop the scan.field case
Likely causes
  • Couch control board not resetting after a power or command glitch
  • Couch control board or encoder fault
Checks, in order
  1. Watch the couch board indicators and review the logs
  2. Check all supply voltages and the CAN bus connector
  3. Check the encoders (blue unit)
  4. Compare the calculated couch travel with the potentiometer feedback
Fix

Probable cause is the couch control board. The recommended steps were to reflash the CPM module on the board and check the supplies and the CAN connector. The case was not confirmed as solved.

Rotor test gives a series of errors after both 25 A fuses on the PI filter board blew. Voltage is present (the neon indicator lights) and the stator windings measure about 2.2 Ω.field case
Likely causes
  • X-ray tube anode failure
  • Rotor spin-up board or soft-start fault
Checks, in order
  1. Confirm supply voltage at the tube housing
  2. Measure the stator winding resistance
  3. Replace the fuses and retest
  4. Check the spin-up board and that the soft start operates
Fix

Probable diagnosis was a failed tube anode. No repair was confirmed.

05

Field notes

Shutdown practice affects the tube and detectors. The guidance passed on was: switch off with the key on the CT box overnight, and do a full power-off only over weekends. With a poor mains supply and no UPS, switching the system off is safer. Tube heat showing 10–14% in the morning was likely because the key switch-off overnight had stopped.

Before going deep into gantry diagnostics on scan aborts, measure the mains voltage under scan load. Low supply voltage caused generator and rotor errors that HSS and rectifier replacement did not fix.

Vicor DC-DC outputs around 3.5–4.0 V on modules labelled 5 V were confirmed as normal by other engineers, so do not replace them on that reading alone.

Before any collimator or RHOST work, take fresh A-plane and NVRAM/host backups. An old backup can differ a lot from the current configuration; after restoring one, check the grid voltage table and redo the filament calibrations.

A humming rotor brake was traced to fine rust dust on the solenoid core preventing full magnetic closure. Cleaning the core (with a very thin lacquer coat on the end faces) was the probable fix.

Repeated pushing of the couch closer than intended bent the couch frame. A floor stop or rail that limits the couch position prevents this.

After any repair, run the manufacturer's full checkout and performance verification before the device goes back to patients.