Early access · courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Library / MRI / MAGNETOM Avanto
Siemens Healthineers

MAGNETOM Avanto

The Siemens Healthineers MAGNETOM Avanto is a superconducting whole-body MRI scanner used for routine and advanced clinical diagnostic imaging.

23 field cases from service engineers' discussions (anonymised); solved: 7, probable: 16under engineering review

02

Faults and how to find them

A PALI error appears now and then on all coils except the knee coil. The RF characterisation test fails at step 4: when the TAS 50-ohm load is switched in, the amplifier goes straight to full power, then slowly drops it, and trips on the first iteration. A related report describes poor or inconsistent image quality that seemed to be linked to tune-up.field case
Likely causes
  • Faulty RFSU power supply (confirmed in one case)
  • Tune-up out of date, or adjustment not being run for each patient (image quality case)
Checks, in order
  1. Read the error log and the test reports before swapping parts
  2. Swap in a known-good TAS to rule it out
  3. Go through the modulator – RFPA – TAS – BCCS – TALES – body coil chain using the results of the first three test steps
  4. Look at the TALES voltages reported by the tests
  5. Check when tune-up was last run and whether adjustment runs for each patient; run TuneUp and Quality, then the RF characterisation tests if the fault remains
Fix

In the PALI / RF test case, replacing the RFSU power supply fixed the fault. For the image quality complaint, the advice was to run TuneUp and Quality first, then the RF characterisation tests; that case had no confirmed outcome.

The Max kV test fails and error 95 appears in the logs. After the RF connectors were checked, the Rx test also started failing.field case
Likely causes
  • A cable on the BDAS that was not fully tightened again after the connector check
Checks, in order
  1. Re-check every RF connector that was disturbed, especially on the BDAS, and make sure each one is fully seated and tightened
  2. Repeat the Max kV test and the Rx test
  3. Scan a phantom to confirm
Fix

Probably a BDAS cable that was not fully tightened after the connector inspection (colleagues' assessment, not confirmed). Re-tightening the connections and repeating the tests is the likely fix.

Coils keep failing. Staff get electric shocks when they touch the RF cage, and static discharges hit the keyboard, mouse and console.field case
Likely causes
  • Static build-up from ordinary linoleum and a synthetic office chair in the control room and the cage, made worse by dry winter air
Checks, in order
  1. Test all coils to see which are actually damaged
  2. Look at the working conditions: floor covering, furniture, clothing, air humidity
Fix

Replacing the linoleum and the chair with antistatic versions solved the problem.

After staff dropped the head coil, a localised image artefact appeared where the coil's metal shield had been bent.field case
Likely causes
  • Bent metal shield inside the coil
Checks, in order
  1. Inspect the coil for mechanical damage after any drop
  2. Compare where the artefact sits in the image with where the coil is damaged
Fix

Straightening the shield solved it in this case. Colleagues doubted that a bent shield alone could cause a local artefact, so treat the cause with caution.

Rollers or bearings in the patient table's vertical (lift/lower) drive have broken up. In one case the lower bearing cap on a guide rail fell apart and some balls fell out. In another, limit switch S30 tripped.field case
Likely causes
  • Worn or broken rollers or bearings on the vertical drive guide rails
  • S30 trips when the table meets an obstacle while moving down (anti-crush protection)
Checks, in order
  1. Check whether something blocked the table on its way down before treating an S30 trip as a fault
  2. Inspect the whole vertical drive while you are in there: bearings, brake, ball screw spindle and nut
  3. Look up the original part numbers in the manufacturer's spare parts catalogue
Fix

Probable fix: replace the rollers or bearings as a complete assembly with their rails. The official procedure needs the magnetic field to be ramped down for access; that must be planned with the manufacturer. Replacing the rail assembly under field was described as possible but extremely difficult.

Scan programs abort when the primary cooling circuit temperature rises. In another case (Avanto Fit), the secondary circuit pump will not start: the pump shows error 42, which is not in the table on the unit cover, pressure is normal, and a green LED is flashing on the pump.field case
Likely causes
  • One of the chiller's two compressors failed, leaving too little cooling capacity
  • Return line pressure too high (above 1.3 bar) on the secondary circuit pump (probable cause of error 42)
Checks, in order
  1. Check the chiller's compressors and the primary circuit temperature trend
  2. Check the heat load in the technical room and the control room, including ventilation
  3. Measure the return line pressure, not just the general circuit pressure
  4. Check the pump's indicator and LED status
Fix

Temporary measure for the chiller case: turning off ventilation in the technical room and control room brought the primary circuit back to a stable 16°C. An extra split air conditioner in the technical room was recommended, and the failed compressor still needs repair. For error 42, high return pressure is the likely cause; no fix was confirmed.

After a power cut, the cold head (F70) compressor takes about 18 minutes to start, and the site calls in worried.field case
Likely causes
  • A built-in start delay, by design: it prevents frequent on/off cycling during power cuts and lets the SEP cabinet get into its operating mode. The secondary circuit pump starts after 3 minutes, which can also make the helium compressor cut out
Checks, in order
  1. Confirm which scanner the compressor is connected to
  2. Check the MSUP signal (DB25), which is involved in controlling the start delay
Fix

This is probably normal behaviour, not a fault. Changing the factory delay is not recommended. Tell operators to wait about 20 minutes after power returns. Colleagues said cycling the MSUP supply breaker starts the compressor at once, but because it involves the cryo system, only do this if the manufacturer's procedure allows it.

Workstation and software issues: clearing the study database takes about 45 minutes and looks like a hang, and it does not reboot by itself. Patient data (for example a wrong year of birth) cannot be edited. In the patient browser the letter 'S' is sometimes lower-case. Patient weight shows in pounds.field case
Likely causes
  • Database clean-up is slow when many studies are still present
  • The study is still open in another task card, or it is protected
  • A lower-case 'S' means only part of the study was transferred
  • Metric units not enabled in configuration
Checks, in order
  1. Do not reboot before the database clean-up has finished; an early reboot left the database uncleared
  2. Close the study in the exam, viewing, filming and 3D task cards before editing
  3. Check whether the study transfer is complete
  4. Look for the 'metric' option in options/configuration; where it is depends on the software version
Fix

Database: delete the studies first, then run AllDBRemove (on Avanto this cut the time from about 35 to 15 minutes), or let it run for about 30 minutes and then reboot manually. That fix was confirmed. Editing data: in the patient browser select the folder, then Edit > Correct; if it is blocked, right-click and use Remove Protect (likely fix). Units: enable 'metric' (likely fix).

The magnet quenches during the first field ramp-up after cool-down or near the final ramp currents, and in some reports with no obvious cause (at night or during a patient scan). Teslameter probe readings during ramp-up were also unreliable.field case
Likely causes
  • Not determined in these cases
  • Dirty contacts on the teslameter probe and the MPS connection (measurement problem during ramp-up)
Checks, in order
  1. Record the helium level before and after the event and the current at which the quench happened
  2. Record the events and hand them to the manufacturer
Fix

No confirmed fix. Ramping, quench investigation and any work on the magnet or helium must be done only by the manufacturer or under its direct guidance. In one ramp-up case, cleaning the probe and MPS contacts with alcohol resolved the measurement problem.

05

Field notes

Before replacing expensive RF parts, read the error log and test reports and swap in known-good modules; one PALI case was caused by the RFSU power supply, not the amplifier.

After inspecting RF connectors, re-check that every cable is fully tightened; a loose BDAS cable is the suspected cause of new Rx test failures.

Repeated coil failures can come from the room, not the scanner: static from flooring and chairs in dry winter air damaged coils until they were replaced with antistatic versions.

Several start-up delays are normal, such as the cold head compressor start after a power cut. Explain this to the site rather than changing factory settings.

Patient table vertical drive bearings are a known wear point. Inspect the whole drive (bearings, brake, spindle, nut) in one go, because full access may need the field ramped down by the manufacturer.

The aspirating smoke detector in the technical room can false-alarm from dust, poor ventilation or a refrigerant leak, but it can also detect a real burning gradient coil or connection; keep it calibrated and tested on schedule. A damaged intercom tube was repaired by shortening and reassembling it, or with 3D-printed parts.

After any repair, run the manufacturer's full checkout and performance verification before the device goes back to patients.