Abbott CELL-DYN Emerald (DxH 500): leak at the wash station that an O-ring change didn't stop
What happened
Dripping from the wash station area; replacing the O-ring did not stop the leak, and on closer inspection the fluid was traced to the RBC measuring chamber, not the wash fitting itself.
What was checked
- The engineers confirmed exactly where the fluid came from — it turned out to be the RBC measuring chamber, not the wash station connection.
- They inspected the measuring electrode wire and its shield where they enter the chamber and found green oxidation along the wire and the epoxy potting.
Cause
Sample fluid had wicked along the measuring electrode wire and its shield through the aged epoxy seal in the RBC chamber; the oxidised joint no longer held, so the drip appeared near the wash station and the O-ring replacement addressed the wrong leak path.
What was done
The electrode wire was replaced: the old epoxy potting was carefully removed with a diamond burr, a new wire was soldered to the electrode and the entry was re-potted with epoxy — delicate workshop work done by experienced personnel. The engineers reported the leak stopped; afterwards the chamber was cleaned and full performance verification per the manufacturer's documentation is required before returning the analyser to use.
Safety
Treat all leaked sample fluid as potentially infectious — wear gloves and eye protection and decontaminate the area before opening anything, and complete the usual electrical safety checks before returning the analyser to service.
Lesson: When a seal replacement does not stop a leak, trace the fluid to its true origin first — in potting-sealed chambers the leak often travels along the electrode wire, not through the O-ring.
Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.