Early access: courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Abbott CELL-DYN Emerald / Micros ES60
Abbott · CELL-DYN Emerald / Micros ES60

Abbott CELL-DYN Emerald / Micros ES60: RBC counts creep upward between cleanings

What happened

A hematology analyzer over-reported RBC parameters; after thorough cleaning the readings returned to normal, but climbed again with each subsequent measurement.

What was checked

Cause

The exact cause was not confirmed — the outcome is probable. Two suspects: contamination of the chambers/insulators that routine cleaning does not fully remove, or a dilution fault — a faulty or leaking valve, a leak in the tubing, or air drawn past the plunger — so the RBC chamber receives less diluent, the sample is effectively more concentrated, and the counts read high.

The cause is probable and was not fully confirmed.

What was done

No confirmed fix in this case. The direction suggested was to exclude the dilution hydraulics first: verify valve operation, leak-check the tubing and the plunger seal, renew worn seals, and continue extended bleach soaking of chambers and insulators as reported in the case, then confirm the result on control material. Deeper work should follow the manufacturer's service documentation and be done by authorised personnel.

Safety

Treat the chambers, insulators and the whole fluid path as potentially infectious material; wear gloves and eye protection when working with bleach, rinse parts thoroughly, and verify the analyzer on control material before it returns to patient samples.

Lesson: When cleaning resets high counts only briefly, suspect not just stubborn contamination but a dilution deficit (valve, tubing, plunger seal) and check how much diluent actually reaches the counting chamber.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes