Abbott · Emerald (гематологический анализатор)

Abbott Emerald: monocytes (Mid) over-reported on third-party diluent

What happened

Reported in the case: an Abbott Emerald hematology analyzer started over-reporting the monocyte/Mid population (15–17%) on every sample after the original diluent was replaced with a third-party reagent from another manufacturer's line. With the original reagents the differential was normal, the blank was clean (zeros) and no error codes appeared.

What was checked

Cause

Most likely (the outcome was probable, not confirmed): the third-party diluent has different osmolality and conductivity than the original, measured cell volumes shift, and part of the lymphocyte/granulocyte population crosses into the Mid gate. The analyzer itself was healthy — it was simply sizing cells against thresholds tuned for the original reagents.

The cause is probable and was not fully confirmed.

What was done

The engineers' suggestion was to re-adjust the differential cell-size threshold (in femtoliters) so the population split matches the reagent in use — a procedure from the manufacturer's service documentation, done by authorised personnel; on some analyzers this menu is not available to regular users. The straightforward alternative is to return to reagents the analyzer is validated for, or use an analogue with documented compatibility and verify the differential on controls before reporting results.

Safety

Treat all patient samples, spilled reagent and analyzer waste as potentially infectious — gloves and eye protection, disinfect spills immediately — and before the analyzer returns to reporting results, confirm a clean blank and passing QC.

Lesson: A hematology differential is only as good as the match between its reagents and its size thresholds, so any switch to a third-party diluent demands blank, control and correlation checks — and often a threshold recalibration — before results can be trusted.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.