ABX Micros 60: vacuum lapped syringe loses seal without grease
What happened
The vacuum lapped (ground-glass) syringe needs periodic greasing: without silicone grease the O-ring seals do not work and the syringe does not hold vacuum.
What was checked
- The engineers checked the syringe and its O-ring seals for dryness and wear, since without grease film the seals are effectively useless.
- They searched for an analogue of the original silicone grease (reported in the case as ABX PN LAM004A) and discussed alternatives, judging them by resistance to alcohol, water and alkali.
- They compared notes with colleagues on which greases have held up in long-term use on this fluidic assembly.
Cause
In this case the lapped syringe depends entirely on a thin film of silicone grease: when the grease dries out or is missing, the O-rings cannot seal and the vacuum path starts to leak.
The cause is probable and was not fully confirmed.
What was done
The engineers renewed the grease on the syringe and seals. The original ABX grease (PN LAM004A, as reported in the case) was the reference; Molykote 111 was reported to hold up for years, and the blue grease used on MEK analyzers was also reported to work and to resist alcohol, water and alkali; dispenser grease was mentioned as another option. The outcome was reported as probable: compatibility with the reagents and sample path should be verified with the manufacturer's documentation before switching to an alternative.
Safety
Fluidic parts of a hematology analyzer carry blood-contaminated samples: wear gloves and eye protection, treat the removed syringe and seals as biohazard and decontaminate them before handling, and confirm no leaks with correct aspiration before returning the analyzer to use.
Lesson: Lapped syringes live on their grease, so keep the correct silicone grease in stock and re-grease on a preventive schedule instead of waiting for a vacuum fault.
Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.