Аккорд · Аккорд 4.0

Accord 4.0: MS SQL 2005 installer fails after OS replacement

What happened

During a deployment of the Acccord 4.0 software on a workstation whose original Windows XP system partition had been wiped and rebuilt with Windows 7 (the installation files were recovered from the second partition), the MS SQL 2005 installer kept aborting with an error. Running it in compatibility mode changed nothing.

What was checked

Cause

The engineers found a localisation problem: the MS SQL 2005 installer configuration was written for an English-language OS, and the service account name it expected ('NT AUTHORITY/system') did not match the account name on a Russian Windows 7 build, so the setup failed.

What was done

The engineers edited the SQL installation file, replacing the 'NT AUTHORITY/system' entries with the Russian localised account name ('NT AUTHORITY/система'); after that the installation completed successfully and the software deployed. The cleaner long-term answer is to rebuild such workstations with the OS version and language the software was supplied for.

Safety

A workstation with a freshly rebuilt OS is returning to clinical use after an intervention: before handover, confirm the clinical application works end-to-end with the connected equipment and, if the maintenance plan requires an electrical safety test after any intervention, perform it.

Lesson: Before rebuilding a workstation, check which OS version and language the application software was certified for — a localized Windows can quietly break old installers that assume English system names.

Covered in the course: DICOM, PACS and Networks — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.