Adani MammoScan: flickering indicator on the lower compression display
What happened
The indicator on the lower display of the compression system flickered intermittently. No fault codes were reported; a poor contact in the compression system was suspected.
What was checked
- As reported in the case: the paddle identification sensor (ID board with a cylindrical Hall sensor) behind the paddle mount was inspected for damage and secure seating.
- As reported in the case: the interface board in the metal box on the left side of the unit was checked, along with its connectors.
Cause
The engineers found a poorly seated ribbon cable in the lower block, where the lower display is located — the intermittent contact made the indicator flicker.
What was done
The ribbon cable was reconnected; the flickering stopped and the unit was returned to use. No parts were needed.
Safety
Before opening the lower block or unplugging connectors, isolate the unit from the mains and make sure the compression paddle cannot move and trap a hand; verify compression operation and run the electrical safety checks before returning the unit to service.
Lesson: When a display indicator flickers, check the simplest thing first — the ribbon-cable connection in the display block — before moving on to sensors and interface boards.