Adani Mammoscan (Pulmoscan series): repeated OnePACS database crashes with loss of studies
What happened
For the second time, as reported in the case, the OnePACS archiving database on the physicians' workstation crashed, and completed studies disappeared from the archive.
What was checked
- Checked the physicians' workstation, since in this setup OnePACS actually stores the studies in a local SQL database there - that is where the archive was failing.
- Ran detector diagnostics as well (noise and supply level) to rule out an acquisition-side cause; reported normal in the case.
Cause
The fault was in the bundled archiving software itself: its SQL database on the workstation proved unstable and kept failing, and each failure took the stored studies with it. The imaging chain was not at fault; the case describes this as a recurring weakness of this archiving setup.
What was done
The engineers replaced OnePACS with Conquest, a free DICOM archive server, on the same workstation and moved study storage to the new archive. As reported, the database failures stopped after the change. Note that swapping the archiving software on a certified system is a configuration change - confirm it is acceptable under your local rules and the manufacturer's support terms before doing the same.
Safety
Before shutting down or restarting the workstation, make sure no exposure can be started, and after the software change verify acquisition, display and archiving with a test exposure before returning the system to clinical use.
Lesson: When a bundled PACS database keeps crashing and losing studies, treat the archiving software itself as the fault - a stable DICOM server in its place stops the data loss, provided such a change is allowed for the device.
Covered in the course: Mammography Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.