Agfa 5302: multiple startup errors after transport
What happened
After the printer was moved to another facility, it produced a burst of errors at every power-on, with no single subsystem clearly at fault.
What was checked
- Shock indicators on the transport packaging were checked first and found intact, so a transit knock was ruled out.
- The rear status LEDs and the power supply were inspected; the engineers then checked the incoming mains lead and found a 3x0.75 cable instead of the required 3x1, with figures as reported in the case.
Cause
When reconnecting the printer after the move, an undersized mains cable had been used; the insufficient conductor cross-section (3x0.75 instead of 3x1, as reported in the case) caused supply instability at power-on that surfaced as a pile of seemingly unrelated errors.
What was done
The undersized cable was replaced with the correct 3x1 mains lead (rating as reported in the case), and the errors no longer appeared at startup; the outcome in this case was a full resolution.
Safety
Before working on the mains connection, isolate the printer from the supply and confirm the earth conductor is intact, and before return to use verify that the cord cross-section and earthing meet the manufacturer's requirements — an undersized cord is a fire and shock risk, not just a source of errors.
Lesson: After any relocation, check the simplest external items first — a mains lead with the wrong cross-section can imitate a whole list of internal faults.
Covered in the course: Medical Imagers and Film Printers — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.