Agfa · CR 10X

Agfa CR 10X: crunching noise and error during cassette erase

What happened

During the erase cycle the digitizer did not pull the imaging plate fully into the cassette; a crunching noise was heard from inside, followed by a fault message.

What was checked

Cause

The engineers found the feed roller dirty: it slipped against the phosphor plate and did not feed it fully into the cassette, so the auxiliary feed ran against the plate — that was the crunching noise — and the unit faulted. The mechanics themselves were undamaged in this case.

What was done

Cleaned the feed roller and the imaging plates (both sides, mild soapy water, well-wrung lint-free cloth), let everything dry fully, reassembled and verified a complete erase cycle. Reported solved in this case.

Safety

Isolate the digitizer from the mains and let the transport mechanism come to a complete stop before removing covers; keep hands clear of rollers and pinch points, and hold plates by the edges to protect the phosphor layer.

Lesson: Slipping, crunching and short feeds in a CR transport usually start with a dirty roller — clean the simple cause before suspecting broken mechanics.

Covered in the course: X-ray Service — Level 2: Fault-Finding →