Agfa CR 10X: crunching noise and error during cassette erase
What happened
During the erase cycle the digitizer did not pull the imaging plate fully into the cassette; a crunching noise was heard from inside, followed by a fault message.
What was checked
- Checked the digitizer transport mechanics for obstruction, wear and free movement.
- Inspected the cassette feed roller; its surface was contaminated and slipping against the plate.
Cause
The engineers found the feed roller dirty: it slipped against the phosphor plate and did not feed it fully into the cassette, so the auxiliary feed ran against the plate — that was the crunching noise — and the unit faulted. The mechanics themselves were undamaged in this case.
What was done
Cleaned the feed roller and the imaging plates (both sides, mild soapy water, well-wrung lint-free cloth), let everything dry fully, reassembled and verified a complete erase cycle. Reported solved in this case.
Safety
Isolate the digitizer from the mains and let the transport mechanism come to a complete stop before removing covers; keep hands clear of rollers and pinch points, and hold plates by the edges to protect the phosphor layer.
Lesson: Slipping, crunching and short feeds in a CR transport usually start with a dirty roller — clean the simple cause before suspecting broken mechanics.
Covered in the course: X-ray Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.