Agfa CR 35x: firmware update blocked by unreadable diskette kit
What happened
During a firmware update the diskette kit was only partly readable and the available disk image threw errors, so the unit kept waiting for a firmware file over the COM port (reported as portex2003.4.zip).
What was checked
- The engineers checked every floppy in the kit and the available image to see what was actually readable.
- They identified from the COM-port requests which firmware file the unit was missing.
- They confirmed the device hard drive could be accessed over SCSI before starting work.
Cause
In this case the cause was degraded floppy media, not the reader: the kit files could not be transferred through the standard path, so the update stalled waiting for the missing file.
What was done
As reported, the engineers obtained a working set of the firmware files, connected the device hard drive to a PC with a SCSI controller, wrote the required files directly onto the HDD, adjusted the update batch scripts to match the actual file set, and ran the update, which completed successfully.
Safety
Before opening the unit to access the drive, isolate it from the mains supply and take ESD precautions with the hard drive, and after the firmware update confirm correct operation with a test exposure before returning the device to clinical use.
Lesson: Verify the update media and have a verified fallback copy of the firmware kit before you start, because a failed update can leave the system down.
Covered in the course: Medical Imagers and Film Printers — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.