Agfa · Drystar 3000

Agfa Drystar 3000: film exits blue, only silhouette printed

What happened

As reported, after about half a day of normal operation the camera began producing film with no image — the sheet came out a uniform blue with only a faint silhouette. Replacing the thermal print head changed nothing.

What was checked

Cause

The film was loaded the wrong way round: the thermal head printed onto the back side, so the image did not transfer and the sheet stayed blue. As reported, the trouble appeared after the film supply had been changed during the day.

What was done

Reloaded the film with the correct side facing the thermal head and ran test prints; image quality, as reported, returned to normal. No parts were needed.

Safety

Isolate the printer from the mains before opening the transport area and let the thermal head cool — it stays hot enough to cause burns, and the unit should pass an electrical safety check before going back into service.

Lesson: When a thermal printer suddenly produces uniform or blank film, check the film side and orientation first — a consumable loaded wrong can perfectly imitate a dead print head.

Covered in the course: Medical Imagers and Film Printers — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.