Agfa · Drystar 5302

Agfa Drystar 5302: film jams when printing from the upper tray

What happened

Films jam during printing, but only when fed from the upper supply tray.

What was checked

Cause

The engineers found that the pickup roller springs had lost tension. The module was nominally new, but as reported in the case it had sat in storage for three years, and the weakened springs let the film slip and jam on pickup.

What was done

As reported in the case, the engineers re-tensioned the pickup springs by four turns and blew out the optical sensors, after which the jamming stopped. They also advise wiping the rubber rollers with alcohol periodically as prevention.

Safety

Disconnect the printer from the mains and let the thermal print head cool down before reaching into the film path.

Lesson: 'New' spare parts age on the shelf: when one tray misfeeds, check pickup roller tension and clean the optical sensors before ordering anything.

Covered in the course: Medical Imagers and Film Printers — Level 2: Fault-Finding →