Agfa Drystar 5302: film type mismatch fault
What happened
The printer reports a film type mismatch: DT2B film is loaded, while the printer requires DT10B.
What was checked
- Read the tag info of the loaded film to identify the film type the printer expects.
- Check the printer's firmware release and confirm which film type that software version requires.
Cause
The engineers found the printer was loaded with DT2B film while, per its firmware release, it requires DT10B — the mismatch triggered the fault.
What was done
Replace the film with DT10B; as reported in the case, DT10B is required on the Agfa 5302 from the third firmware release onward. With the correct film loaded, the printer worked normally.
Safety
Before returning the printer to clinical use, run a test print and verify image density, and if the film transport area was accessed, let hot components cool and refit all covers.
Lesson: Always verify the consumable type against the unit's firmware requirements before loading, since a film mismatch can present as a printer fault.
Covered in the course: Medical Imagers and Film Printers — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.