AGFA DRYSTAR 5302: small squares printed on large films instead of the image
What happened
On large film sizes the printer produced a pattern of small squares instead of a full image. No error codes were reported.
What was checked
- Checked the printer software first — nothing wrong was found there.
- Checked the film type and origin: original Agfa film (DT 2 B) in an original box versus grey-market film, often supplied in re-labelled boxes.
Cause
The engineers found that the film loaded in the printer was not original: grey-market film is not compatible with the DRYSTAR 5302. Reportedly, such film can be adapted for the 5300 model, but no adaptation works on the 5302, so the printer fails to image large films correctly.
What was done
Load original film compatible with the DRYSTAR 5302. No software change was made, and none would have helped — printing returned to normal once the correct film was used.
Safety
After any work inside the imager, verify protective earthing and perform the electrical safety checks required for medical devices before returning the printer to clinical use.
Lesson: For print-quality faults, rule out consumables first — incompatible or grey-market film can fully mimic an electronics or software failure.
Covered in the course: Medical Imagers and Film Printers — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.