Agfa DryStar 5302: two films printed from one job
What happened
Every time a single job was sent, the printer output two identical films. The printer itself showed no faults.
What was checked
- Clarified whether both films came out immediately one after the other from a single send — an immediate pair points to a doubled command, while a delayed second copy would suggest something in the print queue or a repeated entry at the sending side.
- Checked the sending workstation rather than the printer, since the printer logged nothing wrong and the duplication followed the job, not the hardware.
Cause
The engineers found that a faulty mouse on the sending workstation was generating an extra click, so the print command for the same job went out twice. The printer simply executed both commands — the fault was upstream of it.
What was done
The mouse on the workstation was replaced and the duplication disappeared. The printer was not opened and no parts were changed; in this case the problem was fully resolved.
Safety
Disconnect the equipment from the mains before any work inside the printer and let heated and moving parts settle, and repeat the electrical safety test before returning the device to clinical use.
Lesson: Before suspecting the printer, check the sending workstation and its ordinary peripherals — a worn-out mouse can duplicate jobs just as reliably as a board fault.
Covered in the course: Medical Imagers and Film Printers — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.