Agfa · Drystar AXYS

Agfa Drystar AXYS: printer does not boot from USB stick

What happened

The imager would not boot when a USB stick was connected. The engineers were trying to load software from USB and at first took the package reported in the case for an update, when it was in fact a full distribution intended for installation from scratch.

What was checked

Cause

The USB stick had been prepared incorrectly: the distribution archive had not been unpacked, so the required files were not in the root of the stick and the printer had nothing valid to boot from.

What was done

Following the manufacturer's installation documentation, the engineers unpacked the distribution archive and copied all its files to the root of a USB stick, then inserted the stick into the printer's USB port. The camera then booted and installed the software normally; no parts were replaced.

Safety

After any software reload, let the printer complete its full start-up self-test and carry out the electrical safety checks required before returning it to clinical use.

Lesson: A USB software load usually fails because of media preparation, not hardware — unpack the distribution and place its files in the root of the stick before suspecting the printer.

Covered in the course: Medical Imagers and Film Printers — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.