Ahmed LH-10-F: output pressure sags during valve switching

What happened

As reported in the case, the output pressure dropped from about 4 bar to 3 and then 2.5 bar each time the valves switched, while the compressor held a steady 8 bar and the valve block inlet received full pressure.

What was checked

Cause

The engineers found a leak from the receiver downstream of the zeolite columns: the stored volume bled off with every switching cycle, so the output sagged even though the compressor and the valve block inlet stayed at full pressure.

What was done

The leak on the receiver was found and repaired; after that, the output pressure reportedly held steady through valve switching, and the case was closed as solved.

Safety

Bleed the pneumatic circuit down to zero and confirm it on a gauge before opening any connection — the receiver stores pressure energy, and every part of an oxygen concentrator must stay clean and free of oil and grease.

Lesson: When output pressure sags only during valve switching, look at the stored-volume side first — a leaking receiver will drain in step with the switching cycle even when every valve tests tight.