Ahmed LH-10-F: output pressure sags during valve switching
What happened
As reported in the case, the output pressure dropped from about 4 bar to 3 and then 2.5 bar each time the valves switched, while the compressor held a steady 8 bar and the valve block inlet received full pressure.
What was checked
- Checked the pressure relief valve — serviceable.
- Checked the zeolite columns for leaks — none found.
- Verified the pressure at the valve block inlet.
- Leak-tested the valves one by one (six, as reported in the case).
Cause
The engineers found a leak from the receiver downstream of the zeolite columns: the stored volume bled off with every switching cycle, so the output sagged even though the compressor and the valve block inlet stayed at full pressure.
What was done
The leak on the receiver was found and repaired; after that, the output pressure reportedly held steady through valve switching, and the case was closed as solved.
Safety
Bleed the pneumatic circuit down to zero and confirm it on a gauge before opening any connection — the receiver stores pressure energy, and every part of an oxygen concentrator must stay clean and free of oil and grease.
Lesson: When output pressure sags only during valve switching, look at the stored-volume side first — a leaking receiver will drain in step with the switching cycle even when every valve tests tight.
Covered in the course: Fault-Finding Method for Medical Equipment →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.