Aloka Alpha 10: broken trackball retaining tabs during cleaning
What happened
The trackball on an ultrasound console had to be removed for cleaning. While opening the panel, the engineers found the retaining tabs already broken by an earlier, forceful cleaning attempt, which complicated the removal.
What was checked
- Disassembled the panel/keyboard to reach the trackball; on this model the whole keyboard assembly has to come apart before the trackball is accessible.
- Released the trackball shaft: as noted in the case, versions with a retaining tab release it from above, while versions without a tab need the foam end-mounts pressed down and the shaft eased out, nudged with tweezers if it sticks.
- Inspected the tabs and mounting points for damage from the earlier attempt before lifting the ball out.
Cause
The broken tabs were traced to a previous careless cleaning: this trackball family comes in versions with and without retaining tabs, and the shaft is removed differently on each — forcing the wrong method snaps the tabs.
What was done
The fix required a full keyboard teardown: as described in the case, four screws and the handles were removed, the whole keyboard assembly came apart, and the trackball was lifted out, cleaned and reinstalled. Reported effort was about two hours on this model, while the same ball on other models comes out through four screws in minutes. The unit was reassembled and worked normally.
Safety
Before opening the console, switch the system off and unplug it from the mains; after reassembly, check that no tools or fasteners are left inside and run the required electrical safety tests before returning the machine to use.
Lesson: Check the correct removal method for the exact model before forcing a trackball or its retainers during cleaning — the wrong shortcut turns a small job into a full console teardown with broken parts.
Covered in the course: Ultrasound Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.