Aloka SSD-1400: no video signal on the BNC outputs

What happened

The video output died completely: no signal on the BNC connectors, and an oscilloscope on the coaxial cable showed silence.

What was checked

Cause

The power supply had failed: a blown fuse took out the ±15 V rails, so the video circuitry had no power. The boards themselves were fine — the dead supply simply looked like a dead video path.

What was done

Replaced the blown fuse and repaired the power supply; the ±15 V rails came back and the video output returned. The boards were reinstalled into the original unit.

Safety

Unplug the unit and let the PSU capacitors discharge before opening it — mains-side parts stay live; after any power supply repair, run earth-bond and leakage-current tests before returning the machine to use.

Lesson: Measure the supply rails before pulling boards — a missing voltage rail can look exactly like a dead video board, and a multimeter costs far less time than a full board swap.

Covered in the course: Ultrasound Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.