Amico Telemedix: X-ray table stuck at 92 degrees of tilt
What happened
The X-ray table had stopped at a reported 92 degrees of tilt and would not drive back into its normal working range with the regular controls; no error codes were reported.
What was checked
- Confirmed on site: the table sat at the reported 92 degrees and normal controls would not drive it back; no error codes shown.
- Found and studied the manufacturer's service documentation for the tilt drive and the travel-limit arrangement before any intervention.
- Sorted the workaround ideas that came up during troubleshooting: anything that was not a documented procedure for authorised personnel was set aside.
Cause
No confirmed root cause was recorded in the case. A table parked at one angle with no drive back fits the motion resting on its travel-limit switch with the drive refusing further travel, but this was not verified in the case.
What was done
The fix was a procedure from the manufacturer's service documentation, done by authorised personnel: the travel limits were released only for the recovery move to bring the table back into its normal range. Afterwards normal operation and the limit switches were verified, and the table was returned to use.
Safety
While travel limits are overridden the table can move past its normal stops, so keep everyone, and above all any patient, well clear of the moving table and its pinch points, and confirm the limit switches again before the table goes back into service.
Lesson: When motion stops on its travel limit, recover the equipment only with the manufacturer's documented override procedure and re-test the limits before it returns to use.
Covered in the course: X-ray Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.