Amico Telemedix: X-ray table stuck at 92 degrees of tilt

What happened

The X-ray table had stopped at a reported 92 degrees of tilt and would not drive back into its normal working range with the regular controls; no error codes were reported.

What was checked

Cause

No confirmed root cause was recorded in the case. A table parked at one angle with no drive back fits the motion resting on its travel-limit switch with the drive refusing further travel, but this was not verified in the case.

What was done

The fix was a procedure from the manufacturer's service documentation, done by authorised personnel: the travel limits were released only for the recovery move to bring the table back into its normal range. Afterwards normal operation and the limit switches were verified, and the table was returned to use.

Safety

While travel limits are overridden the table can move past its normal stops, so keep everyone, and above all any patient, well clear of the moving table and its pinch points, and confirm the limit switches again before the table goes back into service.

Lesson: When motion stops on its travel limit, recover the equipment only with the manufacturer's documented override procedure and re-test the limits before it returns to use.