Amiko Optimax: developer and fixer supply and drain hookup

What happened

A question about the correct connection of the developer and fixer supply, recirculation and drain lines; the initial assumption was that both solutions return (recirculate) back into their tanks.

What was checked

Cause

The initial assumption that the solutions recirculate back into their tanks was wrong for this machine.

What was done

Reported in the case: the developer is routed straight to drain without any recirculation loop, because returning it to the tank degrades development quality; the fixer is drained through a valved drain with an overflow and collected for disposal, in line with local waste rules. With this hookup the issue was resolved.

Safety

Before re-routing any chemistry lines, isolate the processor from the power supply and work in chemical-resistant gloves and eye protection; all spent developer and fixer must be handed over for disposal according to local hazardous-waste rules, never simply poured away.

Lesson: Always verify the manufacturer's plumbing scheme before hooking up chemistry lines — an assumed recirculation loop can quietly degrade image quality.