Early access · courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Аполло МТЛ
Аполло · МТЛ

Apollo MTL: collimator lamp replacement and switching relay failures

What happened

The collimator alignment lamp had to be replaced and its correct rating confirmed; overrated lamps were reported to burn out the relay in the cabinet that switches the lamp.

What was checked

Cause

In this case, lamps rated 150 W (as reported in the case) drew more current than the lamp-switching relay in the cabinet was designed for, so the relay burned out; the rating confirmed by the documentation was 24 V, 100 W.

What was done

The engineers fitted the lamp rating confirmed by the documentation — 24 V, 100 W as reported in the case — instead of the 150 W type. At the lower rating the relay holds, the lamp runs steadily, and the light is sufficient for collimator alignment. The outcome was reported as solved.

Safety

Isolate the unit from the mains and let the lamp and housing cool before opening the collimator, and after any work in the X-ray room complete the manufacturer's electrical and radiation checks before return to use.

Lesson: Fit only the lamp rating given in the unit's documentation — a higher-wattage lamp may appear to work, but it can quietly overload and destroy the relay that switches it.

Covered in the course: X-ray Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes