ARMED 7F-10L: shuts down about a minute after start

What happened

As reported in the case, the concentrator shut itself off roughly a minute after startup, but with the nebulizer valve open it kept running normally. No error codes were shown.

What was checked

Cause

The engineers found that the fault was tied to one of the columns: after the columns were swapped the machine ran normally. The exact component or mechanism was never identified, so the cause should be treated as probable, not proven.

What was done

The columns were swapped between positions one at a time; after that the unit ran stably, with the oxygen concentration reported at 94%. No parts were replaced — this was a fault isolation by substitution, not a repair.

Safety

Before returning an oxygen concentrator to use, verify the delivered oxygen concentration and complete the electrical safety test, and keep the device away from oils, greases and open flame because of the oxygen-enrichment fire risk.

Lesson: When a fault follows no obvious electrical pattern, swapping identical assemblies one at a time remains the fastest way to pin it down.