Armed LF H 10l: valve disc wear and consumable replacement

What happened

No active fault: routine service review for the concentrator — which consumables fail first and how often to replace them, with worn valve rubber discs reported as the most common failure in this case.

What was checked

Cause

No single fault: ordinary wear of consumable parts — valve rubber discs lose elasticity, zeolite degrades faster in humid rooms, moisture separator and filters saturate and clog over time. The engineers found this in one case, not as a rule.

What was done

Scheduled preventive replacement of wear items, as reported in the case: valve rubber disc kit (12 discs), zeolite every 5–6 years depending on room humidity, moisture separator (internals or complete), air and compressor filters, sealing rings, and high-pressure tubing when worn. Also noted: the concentrator's output is not suitable for calibrating the oxygen sensor of some ventilators — perform such calibration with the setup specified in the ventilator manufacturer's service documentation.

Safety

Unplug the concentrator before opening it, keep oil and grease away from oxygen-path parts because oxygen-enriched materials ignite easily, and run an electrical safety test before return to use.

Lesson: On pressure-swing oxygen concentrators, treat small wear parts — valve rubber discs, filters, moisture separator, zeolite — as scheduled consumables, because oxygen purity drops gradually before the unit gives any clear sign.

Covered in the course: Fault-Finding Method for Medical Equipment →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.