Armed SN211-130 (Recirculator 2-130): UV lamps and ballast board burning out
What happened
On several units of this model (three of four, as reported in the case) the UV lamps failed and the ballast board burned out: the engineers found a blown fuse and a burnt trace on the board.
What was checked
- Inspect the ballast board: check the fuse for continuity and look for burnt or open traces.
- Test the UV lamps themselves — in this case they turned out to be the source of the overload.
- After any ballast replacement, check transistor temperature on the old lamps: if they run hot, the lamps are the problem, not the new ballast.
Cause
The engineers found that the lamps themselves were faulty (poor quality) and drew abnormal current, overloading the ballast board until the fuse blew and a trace burned out. This is the conclusion from one case, not a general rule for the model.
What was done
Restore the burnt trace, fit a new fuse, and replace all UV lamps with good-quality ones. In one unit the whole ballast was swapped, but the transistors still ran hot on the old lamps — replacing the lamps cured it. Per the case, the units then ran without problems for about six months.
Safety
Isolate the unit from the mains and let the ballast capacitors discharge before opening it, never work with UV-C lamps powered, and before returning the unit to use run an electrical safety check and confirm the enclosure fully shields the UV lamp.
Lesson: When the same failure repeats across identical units, suspect the lamps before condemning the ballast — a new ballast run on the old lamps can fail the same way.
Covered in the course: Steriliser Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.