Armed SN211-130 (Recirculator 2-130): UV lamps and ballast board burning out

What happened

On several units of this model (three of four, as reported in the case) the UV lamps failed and the ballast board burned out: the engineers found a blown fuse and a burnt trace on the board.

What was checked

Cause

The engineers found that the lamps themselves were faulty (poor quality) and drew abnormal current, overloading the ballast board until the fuse blew and a trace burned out. This is the conclusion from one case, not a general rule for the model.

What was done

Restore the burnt trace, fit a new fuse, and replace all UV lamps with good-quality ones. In one unit the whole ballast was swapped, but the transistors still ran hot on the old lamps — replacing the lamps cured it. Per the case, the units then ran without problems for about six months.

Safety

Isolate the unit from the mains and let the ballast capacitors discharge before opening it, never work with UV-C lamps powered, and before returning the unit to use run an electrical safety check and confirm the enclosure fully shields the UV lamp.

Lesson: When the same failure repeats across identical units, suspect the lamps before condemning the ballast — a new ballast run on the old lamps can fail the same way.

Covered in the course: Steriliser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.