Early access · courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Avent Avent M
Avent · Avent M

Avent M: turbine will not start, calibrations fail

What happened

The ventilator would not spin up the turbine and failed its startup calibrations — after power-on it essentially did nothing. No error codes were reported in the case.

What was checked

Cause

The engineers found that the turbine control board had burned out, most probably because of a shorted (punctured) capacitor. It was noted in the case that on turbines with a transparent bottom such failures can be rather spectacular.

What was done

The turbine control board was replaced or repaired; after that the turbine started and the calibrations passed again. Any post-replacement setup and calibration were done with procedures from the manufacturer's service documentation by authorised personnel.

Safety

A burned board can hold charge in its capacitors and can smoulder after power-off: disconnect the unit from the mains, let the capacitors discharge and inspect for heat and smoke damage before touching anything, and complete the electrical safety test and calibration before the ventilator returns to patient use.

Lesson: When a turbine ventilator does nothing at startup, check the turbine control board indication and identify the turbine type early — a dead board hides behind generic calibration failures.

Covered in the course: Ventilator Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes