Early access · courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Awareness Technology Stat Fax 1904
Awareness Technology · Stat Fax 1904

Awareness Technology Stat Fax 1904: user methods lost after battery replacement

What happened

After the internal battery was replaced, the stored user methods (adaptations) were gone, and no new method could be created and no existing one edited.

What was checked

Cause

In this case the engineers attributed the loss to the custom methods being kept in battery-backed memory: once the battery was removed, the settings were wiped. The locked first positions were a design feature of the instrument, not a fault.

What was done

The parameters were re-entered by hand from the reagent inserts: wavelength, measurement method and calibrator concentration. Since the first positions are only partly locked and method names cannot be assigned, the settings were written in the order of the test numbers, using the printed list of names as a reference, and saved. After that, the engineers could create and edit methods again, and the case was reported as solved.

Safety

Isolate the analyzer from the mains before opening the housing to replace the internal battery, and verify the restored methods with control material before returning the instrument to use.

Lesson: Keep an off-instrument copy of every user-defined method, so that a battery-backed memory loss can be rebuilt from documentation instead of from memory.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes