Awareness Technology Stat Fax 2100: unknown square element under the lamp
What happened
An engineer asked what the square element under the lamp is for; no fault symptoms were reported in the case.
What was checked
- The engineers found that the square element is a plain sticker in the light path, not damage or a leftover part.
- Reported in the case: it is a piece of the same tape as the strips on both sides under the carriage.
- Reported in the case: the element works as a light attenuator and is used when readings of the built-in lamp test (test 186) go above the upper limit of the range, roughly 2–10 V per the service manual.
Cause
No fault: the square element is an intentional light attenuator that reduces lamp output when the detector voltage would otherwise rise above the upper calibration limit.
What was done
No repair was needed. The engineers explained that the element is a piece of the same tape as the strips on both sides under the carriage; fitting or trimming such an attenuator is done according to the lamp-output test readings and the manufacturer's service documentation, and remains a task for authorised personnel.
Safety
Before reaching into the optical bay, disconnect the analyzer from the mains and let the lamp cool; the lamp and its housing stay hot enough to cause burns, and avoid looking directly into the beam when the lamp is on.
Lesson: In a photometer, small tapes and filters in the light path are deliberate calibration aids, so check the service documentation before removing anything that looks out of place.
Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.