B.Braun Compact S: time-to-service counter needs correction
What happened
After work on the pump, the time-to-service counter stored in device memory had to be reset or corrected, otherwise the pump would keep indicating that service is due.
What was checked
- The engineers reported reading the pump's non-volatile memory to obtain a full dump of its contents.
- They compared dumps taken after different periods of operation to locate the field holding the service counter and to determine the size of that field, without any confirmed result at the time of the case.
Cause
No separate hardware fault was identified in this case: the need came from the stored time-to-service value itself, which must be set correctly after maintenance according to the manufacturer's documentation.
The cause is probable and was not fully confirmed.
What was done
The counter correction is a procedure from the manufacturer's service documentation, done by authorised personnel. The direct memory-dump editing discussed in the case is not a documented fix, and the outcome was only probable — no verified result was reported.
Safety
Before the pump returns to clinical use, run its full self-test, verify occlusion and air-in-line alarms, and complete an electrical safety test, since the service counter supports preventive maintenance that keeps infusion therapy safe.
Lesson: Service counters exist to enforce maintenance intervals, so reset them only through the manufacturer's documented procedures — editing memory by hand risks corrupting device data and erasing the record that maintenance was due and actually done.
Covered in the course: Fault-Finding Method for Medical Equipment →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.