B. Braun Perfusor Compact: expired service interval
What happened
The pump reported that its service time interval had expired and requested a service interval reset; in the reported case it was otherwise working normally.
What was checked
- The engineers found the prompt was the routine due-for-service flag, not a fault, and the pump otherwise passed its everyday function checks.
- They confirmed the official reset needs the manufacturer's dedicated service cable and software, so it can only be done by authorised personnel.
Cause
The accumulated operating time had simply reached the end of the planned maintenance interval — a scheduled-service reminder, not a hardware failure.
The cause is probable and was not fully confirmed.
What was done
Because the reset depends on manufacturer-only tooling, the fix was a procedure from the manufacturer's service documentation, done by authorised personnel: scheduled service with calibration verification and the interval counter reset. The unofficial firmware workaround discussed in the case was ruled out — such edits put calibration integrity and traceability at risk. The reported outcome was probable, not confirmed: the pump was expected back in use after the documented service.
Safety
Before the pump returns to clinical use, run the manufacturer's functional and flow-accuracy verification plus an electrical safety test — on a drug-delivery device the service interval is a patient-safety reminder and must never be bypassed.
Lesson: A service-interval lock enforces safety-critical maintenance, so the answer is documented service by authorised personnel, never counter or firmware tampering.
Covered in the course: Fault-Finding Method for Medical Equipment →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.