B. Braun Perfusor Compact Plus: fault message at power-on

What happened

The syringe pump displayed a fault message at power-on (reported in the case as error 22) and did not reach ready state.

What was checked

Cause

The engineers found a broken conductor in the keyboard ribbon cable, so at start-up the controller detected a faulty keypad.

What was done

The keypad was replaced as an assembly; after replacement the pump powered up normally and the reported error did not return.

Safety

Before returning the pump to service, perform the electrical safety test and a basic delivery functional check as required by the manufacturer's service documentation.

Lesson: A fault message at start-up is not always the electronics — the keypad or its flat cable can be the culprit, and a quick substitution test with a known-good part is often the fastest way to confirm it.

Covered in the course: Fault-Finding Method for Medical Equipment →