B. Braun Perfusor Compact Plus: fault message at power-on
What happened
The syringe pump displayed a fault message at power-on (reported in the case as error 22) and did not reach ready state.
What was checked
- The engineers first checked what the reported error code means for this model; in this case it pointed to the keypad and its flat cable.
- As a substitution test, the keypad from a working unit of the same model was fitted; the fault disappeared with the substitute keypad, confirming the keypad itself.
Cause
The engineers found a broken conductor in the keyboard ribbon cable, so at start-up the controller detected a faulty keypad.
What was done
The keypad was replaced as an assembly; after replacement the pump powered up normally and the reported error did not return.
Safety
Before returning the pump to service, perform the electrical safety test and a basic delivery functional check as required by the manufacturer's service documentation.
Lesson: A fault message at start-up is not always the electronics — the keypad or its flat cable can be the culprit, and a quick substitution test with a known-good part is often the fastest way to confirm it.