B. Braun Perfusor Compact Plus: syringe pump refuses to start at power-on
What happened
Reported in the case: error 0750 appears at switch-on and the syringe pump does not allow infusion to be started.
What was checked
- At switch-on, watch for any indication that a key is held pressed; run through all membrane keys and feel for one that sits low or does not return.
- Inspect the keypad for dried infusion fluid, disinfectant residue or a dented key; clean the front panel gently and re-test before opening the housing.
Cause
The engineers considered the most likely cause to be one sticking membrane key: the pump detects a pressed key during the power-on self-test and blocks start-up until the keypress clears. In this case the attribution was probable, not confirmed by a full teardown.
The cause is probable and was not fully confirmed.
What was done
The fix in this case was to free the sticking key: clean around it, restore normal key travel and re-check power-on; if the membrane keypad does not recover, it is replaced as a spare part following the manufacturer's service documentation.
Safety
Before any keypad work, disconnect the pump from mains power and remove the syringe with the giving set, and before the pump returns to the ward confirm that the start-up self-test, audible alarms and occlusion alarm pass, so the device is demonstrably safe for drug delivery.
Lesson: A start-up error that appears before any infusion is often the self-test catching a stuck keypad key, so check the keys first before suspecting the electronics.
Covered in the course: Fault-Finding Method for Medical Equipment →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.