B. Braun · Perfusor compact S

B. Braun Perfusor compact S: fault alarm at power-on after battery removed while running

What happened

At power-on the syringe pump shows a fault alarm, reported in the case as error 022, instead of starting up normally.

What was checked

Cause

Improper shutdown: mains was removed without the power control and the battery was pulled from the still-running pump, so the battery monitoring flagged a fault on power-up.

What was done

Power the pump off with the ON/OFF control and keep it held until the alarm symbol clears; if a further alarm appears afterwards (reported in the case as error 105), leave the unit switched off until it is checked. Correct shutdown order: power off via the control first, then remove the cord or the battery.

Safety

Never pull the mains cord or the battery on a running infusion pump — it can interrupt therapy and leaves the unit in an unverified state; shut it down with the power control, disconnect from mains before handling the battery pack, and confirm the full start-up self-test before the pump is returned to use.

Lesson: An alarm right after an unusual power-down is often just the device reacting to how it was switched off — ask how the unit was shut down before suspecting the battery or hardware.

Covered in the course: Fault-Finding Method for Medical Equipment →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.