B. Braun Perfusor compact S: fault alarm at power-on after battery removed while running
What happened
At power-on the syringe pump shows a fault alarm, reported in the case as error 022, instead of starting up normally.
What was checked
- The engineers found that the unit had been disconnected from mains by pulling the power cord rather than with the power button, so the pump kept running on battery.
- Because of that, the battery pack was removed while the unit was still switched on, which triggered the battery alarm at the next start-up.
Cause
Improper shutdown: mains was removed without the power control and the battery was pulled from the still-running pump, so the battery monitoring flagged a fault on power-up.
What was done
Power the pump off with the ON/OFF control and keep it held until the alarm symbol clears; if a further alarm appears afterwards (reported in the case as error 105), leave the unit switched off until it is checked. Correct shutdown order: power off via the control first, then remove the cord or the battery.
Safety
Never pull the mains cord or the battery on a running infusion pump — it can interrupt therapy and leaves the unit in an unverified state; shut it down with the power control, disconnect from mains before handling the battery pack, and confirm the full start-up self-test before the pump is returned to use.
Lesson: An alarm right after an unusual power-down is often just the device reacting to how it was switched off — ask how the unit was shut down before suspecting the battery or hardware.
Covered in the course: Fault-Finding Method for Medical Equipment →