BD FACSCount: low laser power alarm not cleared by prime and long rinse
What happened
The analyser flagged low laser power. A system prime and a full (long) rinse changed nothing, so air in the fluidics was largely ruled out early.
What was checked
- System prime and a full (long) rinse to rule out air bubbles or poor sample flow masking the reading.
- Inspection of the laser power supply and verification of its output before blaming the tube.
- Any adjustment of the laser drive current was treated as a procedure from the manufacturer's service documentation, done by authorised personnel — not as a quick turn of a trimmer.
Cause
Probable, as reported in the case: degradation of the laser itself — the output has fallen below the level the instrument accepts, while the power supply was still assumed healthy. The engineers found that prime and full rinsing had no effect, which points to the light source rather than the fluidics.
The cause is probable and was not fully confirmed.
What was done
In this case the proposed path was: confirm the power supply output first; if the supply is within specification and the laser still reads low, replace the laser and realign. Replacement and alignment are a procedure from the manufacturer's service documentation, done by authorised personnel. Raising the drive current on an ageing tube was not accepted as a fix in the discussion — it only postpones replacement and can finish the tube off faster. The outcome was probable, not confirmed by a repaired unit.
Safety
Laser power supplies hold lethal stored charge and the beam may be invisible: isolate the analyser from mains and allow capacitors to discharge before opening anything, never defeat interlocks, and run the electrical safety test before return to use.
Lesson: When a laser-power alarm survives prime and a full rinse, stop chasing the fluidics and verify the light source and its supply — treating any current adjustment as a documented job, not a quick trim.
Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.