Beckman Coulter · Access 2

Beckman Coulter Access 2: white slime in drain line after allergen runs

What happened

White mucus-like slime builds up in the drain line and collector feeding the waste canister, appearing after allergen test runs. Over time the slime accumulates and the dilution block clogs.

What was checked

Cause

Most probably residue from allergen assays settling in the drain path and building up into slime; the exact composition of the deposits was not confirmed in the case.

What was done

Cleaned as reported in the case: chlorine gel applied in its channel; the substrate channel washed with Citranox at a 1:10 dilution, then with 70% alcohol; the water system flushed with a 2% alkaline solution. The drains were cleared with a large Janet-type syringe with tubing, pushed through the opening of the removed tower after taking off its valve. The engineers also periodically remove the dilution block — three bolts, imperial Hex keys — and clean underneath it. After this the line ran clear, as reported in the case.

Safety

Treat all fluid paths, drains and the waste canister as potentially infectious; wear gloves and eye protection when handling chlorine and alkaline solutions, and confirm a clean rinse with no leaks before returning the analyzer to use.

Lesson: Slime in drain lines is easier to prevent than to clear, so periodic disassembly and cleaning of the dilution block and collector belongs in the planned maintenance routine, not in the response to a clog.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.