Beckman Coulter DxC700 AU: RO water treatment undersized, insufficient output
What happened
The analyzer's documentation calls for 20 L/h of purified water, but the proposed water treatment relied on small reverse-osmosis membranes, so there was doubt it could hold the required flow; the small membranes also put the quality of their rinsing in question.
What was checked
- Confirm the required purified-water flow against the analyzer documentation — 20 L/h in this case.
- Count the RO membranes and check their surface area against the required flow.
- Check the deionization resin volume (fill volume, cylinder size).
Cause
The engineers found an undersized water treatment configuration: small RO membranes were selected instead of a properly sized system, so it could not guarantee the analyzer's flow and water-quality needs.
What was done
Rebuild the water treatment: re-plumb the six membranes into two parallel branches of three each and use deionization resin in 21-inch cylinders. In a reported example on the same analyzer model, a 4021 membrane with two bags of resin held excellent water quality for five months; this case was reported as solved.
Safety
Isolate and depressurize the water lines before re-plumbing them, and before returning the analyzer to use verify the purified water (resistivity and bacterial quality) against the manufacturer's requirements.
Lesson: Size the water treatment from the analyzer's documented flow and purity requirements, not from what is easiest to supply — undersized membranes and resin resurface later as water-quality and reagent problems.
Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.