Beckman Coulter · DxH 600/800

Beckman Coulter DxH 600/800: fluidic buildup from skipped overnight washes

What happened

Risk of background and sample carryover in the fluidic path when the overnight wash and daily shutdown are skipped or done with the wrong agent; the case reviewed the maintenance routine that keeps the analyzer clean.

What was checked

Cause

Protein and reagent residue accumulating in the fluidic path and on the probe when the scheduled overnight wash and shutdown cleaning are missed or replaced with the wrong chemistry.

What was done

The engineers reported the routine that worked: overnight bleach fill at the dilution reported in the case, daily shutdown with the enzymatic cleaner (never bleach for that step), and only occasional vinegar decontamination. Concentrations and sequences were taken from the manufacturer's service documentation and performed by trained personnel.

Safety

Never mix chlorine bleach with vinegar or other acids — the reaction releases toxic chlorine gas; flush the fluidics thoroughly between agents, wear eye and hand protection, and treat all fluidic waste as potentially infectious.

Lesson: Cleaning chemistry is sequence-specific: enzymatic cleaner for the daily shutdown, bleach for the overnight wash, acid only occasionally — never mixed, and always at the documented dilution.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.