Bio-Rad PW40: language and wash settings reset after power failure
What happened
After a mains power failure the microplate washer comes back with the interface in English and the programmed wash routines reset; a language change holds only until the next reboot.
What was checked
- Checked whether the analyzer carries any backup cell at all — there was a firm belief that the unit had no battery inside.
- Confirmed that the resets repeat after every power cycle, meaning retained settings are lost rather than a one-off glitch.
Cause
The engineers found a discharged timekeeper backup cell — with it dead, the analyzer cannot hold language and wash settings across a power cycle.
The cause is probable and was not fully confirmed.
What was done
Replace the timekeeper cell — described in the case as a yellow cell that reportedly gets renewed about once in ten years — then re-enter the settings. The outcome in this case was probable; final confirmation would come after the next unplanned power loss.
Safety
Isolate the analyzer from the mains and wait for internal capacitors to discharge before opening the housing, and complete an electrical safety test before returning the washer to routine use.
Lesson: When a device loses settings at every power-up, check the timekeeper backup cell first — the belief that