Early access: courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Bio-Rad PW40
Bio-Rad · PW40

Bio-Rad PW40: language and wash settings reset after power failure

What happened

After a mains power failure the microplate washer comes back with the interface in English and the programmed wash routines reset; a language change holds only until the next reboot.

What was checked

Cause

The engineers found a discharged timekeeper backup cell — with it dead, the analyzer cannot hold language and wash settings across a power cycle.

The cause is probable and was not fully confirmed.

What was done

Replace the timekeeper cell — described in the case as a yellow cell that reportedly gets renewed about once in ten years — then re-enter the settings. The outcome in this case was probable; final confirmation would come after the next unplanned power loss.

Safety

Isolate the analyzer from the mains and wait for internal capacitors to discharge before opening the housing, and complete an electrical safety test before returning the washer to routine use.

Lesson: When a device loses settings at every power-up, check the timekeeper backup cell first — the belief that

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes