BIOBASE · BKM-Z60B

BIOBASE BKM-Z60B: random valve and vacuum pump activation with recurring faults

What happened

The autoclave intermittently flagged faults, and valves and the vacuum pump switched on and off on their own between cycles.

What was checked

Cause

Water was seeping along the temperature sensor onto the control board, producing random inputs and chaotic switching of outputs.

What was done

The engineers stopped the leak by resealing the temperature sensor fitting and dried the board; the random switching stopped. Reported as solved in this case.

Safety

Before opening a sterilizer, fully depressurize it and let it cool, since hot water and steam store serious energy, and because liquid reached the control board, run electrical safety checks (earth continuity, leakage/insulation) before the unit goes back into use.

Lesson: Random, chaotic switching in a sterilizer means check for moisture on the electronics first — trace every wet path, including sensor ports and fittings, before replacing components.

Covered in the course: Steriliser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.