BIOBASE BKM-Z60B: random valve and vacuum pump activation with recurring faults
What happened
The autoclave intermittently flagged faults, and valves and the vacuum pump switched on and off on their own between cycles.
What was checked
- As reported in the case, the engineers checked and replaced the likely components one by one — door lock, valves, heating elements — but the fault kept returning.
- With the failure staying random, they inspected the control board and found water dripping onto it through the temperature sensor; its fitting and mounting were letting water through.
Cause
Water was seeping along the temperature sensor onto the control board, producing random inputs and chaotic switching of outputs.
What was done
The engineers stopped the leak by resealing the temperature sensor fitting and dried the board; the random switching stopped. Reported as solved in this case.
Safety
Before opening a sterilizer, fully depressurize it and let it cool, since hot water and steam store serious energy, and because liquid reached the control board, run electrical safety checks (earth continuity, leakage/insulation) before the unit goes back into use.
Lesson: Random, chaotic switching in a sterilizer means check for moisture on the electronics first — trace every wet path, including sensor ports and fittings, before replacing components.