BioSystems SA1500: Intermittent probe positioning error
What happened
The probe of the sampling assembly periodically failed to reach its programmed positions. The error was reported as independent of reagent level and appeared even with full volumes, so fluid level was not the trigger.
What was checked
- Compared behaviour at different reagent levels, including full volume — the error still appeared, so reagent level was ruled out as a factor (as reported in the case).
- Adjusted the probe on the X/Y/Z axes — the intermittent errors persisted, so simple mechanical alignment was not the whole story.
Cause
Probable cause: the stored position calibration of the probe axes had drifted or been set incorrectly, so the drives went to slightly wrong coordinates. This conclusion is drawn from one similar case, not a general rule, and was not confirmed by deeper diagnostics or parts replacement.
The cause is probable and was not fully confirmed.
What was done
On a similar SA1500 the symptom was cleared by a manual (not automatic) calibration of all stored probe positions — a procedure from the manufacturer's service documentation, done by authorised personnel. For the discussed case the outcome was considered probable: automatic calibration alone had not stopped the errors, and axis adjustment had only reduced their frequency.
Safety
Treat the probe, fluid path and any spilled sample or reagent as biohazardous, de-energise the analyzer before working near the moving axes, and complete the required QC and electrical safety checks before returning the unit to samples.
Lesson: If a positioning error does not depend on volume or load, suspect the stored position calibration and verify every position with the manufacturer's documented manual procedure instead of relying on the automatic routine.
Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.