Boule Swelab: DP error on hydraulic fill
What happened
The analyzer stopped with a DP fault. Service documentation lists many possible causes for this error, and the deciding factor is the fill time of the second, thicker tube.
What was checked
- Read the manufacturer's service documentation: the DP error has many possible causes, identified by the fill time of the second (thicker) tube.
- Watched the fill of that tube and traced the fluid path through the valves.
- Inspected the valve tubing and found a hairline crack in one of the tubes.
Cause
The engineers found a micro-crack in one of the valve tubes; the leak disturbed the fill time and triggered the DP error.
What was done
The cracked tube was replaced with a new one; after that the analyzer ran without the DP error.
Safety
Treat the fluid path and any spilled sample as potentially infectious: wear gloves and eye protection, disinfect spills at once, and complete the recommended cleaning and a functional check before returning the analyzer to use.
Lesson: A hairline crack in ordinary tubing can mimic a complex fault, so check the wear parts against the documented fill timings before blaming anything electronic.