Boule · Swelab

Boule Swelab: DP error on hydraulic fill

What happened

The analyzer stopped with a DP fault. Service documentation lists many possible causes for this error, and the deciding factor is the fill time of the second, thicker tube.

What was checked

Cause

The engineers found a micro-crack in one of the valve tubes; the leak disturbed the fill time and triggered the DP error.

What was done

The cracked tube was replaced with a new one; after that the analyzer ran without the DP error.

Safety

Treat the fluid path and any spilled sample as potentially infectious: wear gloves and eye protection, disinfect spills at once, and complete the recommended cleaning and a functional check before returning the analyzer to use.

Lesson: A hairline crack in ordinary tubing can mimic a complex fault, so check the wear parts against the documented fill timings before blaming anything electronic.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →