Breas Vivo: battery charge cycle count not accessible
What happened
An engineer needed to read the battery charge cycle counter on a Breas Vivo ventilator but could not find the value in the routine menus; the question was how to reach the level where that counter is stored.
What was checked
- In the reported case, the engineers first checked whether battery information appears in the standard user or clinician-level menus; on this ventilator it did not, at least not in full.
- They then checked with colleagues and against the manufacturer's documentation where the counter actually sits — behind protected menu access, with detailed counters and logs available only through the manufacturer's PC software connected by service cable.
Cause
No hardware fault was involved. The counter simply sits in protected service data that is not shown in day-to-day menus, so the blocker was access, not a failed component.
The cause is probable and was not fully confirmed.
What was done
As reported in the case, the probable solution was a procedure from the manufacturer's service documentation, done by authorised personnel: battery statistics are read through the device's protected service access, and detailed counters and logs through the manufacturer's PC software connected with the service cable. No parts were replaced.
Safety
Before the ventilator returns to a patient, run its full pre-use self-test and verify delivered ventilation on a test lung; a battery with a high cycle count may hold less backup reserve, so confirm the power-fail alarms and battery switchover behaviour as well.
Lesson: Battery condition data on life-support ventilators usually sits behind protected service access, so plan battery assessment around the manufacturer's authorised route and documentation rather than hunting for undocumented menu entry.
Covered in the course: Ventilator Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.