Canon Aplio: image artifacts on one transducer only
What happened
On a Canon Aplio ultrasound, interference artifacts appeared in the image on one transducer only. The engineers suspected mains power quality rather than the probe.
What was checked
- Try a different wall outlet and run the scanner with the UPS out of the circuit — see if the artifacts change (reported in the case).
- Confirm the artifacts are tied to the one transducer: plug it into another connector and try another transducer, to separate a probe fault from mains interference.
- Reverse the mains plug as a quick polarity test (where the plug allows it) and check the condition of the outlet and its earthing.
- Move the scanner to another room / another supply circuit and compare the image.
- Note: repairs on the mains side and electrical safety verification must follow the manufacturer's service documentation, done by authorised personnel.
Cause
The engineers found that the interference came from noisy mains power at that particular outlet — it entered through the supply and showed up on the one transducer, not from a faulty probe or board (reported in the case).
What was done
In this case moving the scanner and plugging it into a different outlet on another supply circuit cleared the artifacts; ferrite rings on the cables were also recommended as extra suppression (reported in the case).
Safety
After any change on the mains side, confirm the scanner is fed from a properly earthed outlet and carry out the electrical safety checks — earth continuity and leakage current — before returning it to clinical use.
Lesson: When artifacts show up on only one transducer, don't condemn the probe straight away — a quick supply-side test (different outlet, UPS out of circuit) can reveal mains interference before you swap expensive parts.
Covered in the course: Ultrasound Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.