Canon CT: scanner freezes and thin-slice reconstructions stop auto-starting

What happened

In this case the CT froze during normal work, and reconstructions of thin slices stopped starting on their own — they had to be launched manually.

What was checked

Cause

The engineers found that the local study database was full. A CT's on-board database is working storage, not an archive; when it overfills, the system stalls and background jobs such as automatic reconstructions stop being created.

What was done

In this case the engineers freed the database by deleting studies from previous shifts, after which the pending thin-slice reconstructions were started manually. The scanner returned to normal operation; no parts and no service-level access were involved.

Safety

Before deleting anything, confirm the studies are archived where the facility expects to find them — deleted patient images are usually unrecoverable and may be needed for care; then verify normal operation with a test acquisition before returning the scanner to clinical use.

Lesson: Treat a CT's local database as working storage that needs active housekeeping — when it fills up, freezes and stalled reconstructions are the typical first signs.

Covered in the course: CT Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.