Canon Go: table control board (TCB) test before replacement

What happened

Suspected fault on the table control board (TCB) of a CT; a spare board had to be prepared and tested before deciding on the swap, and the drive units (VFDs) were also bench-checked.

What was checked

Cause

No confirmed root cause was reported: the original TCB passed testing and was returned to service, so its failure was not proven.

What was done

The replacement was done as a procedure from the manufacturer's service documentation, performed by authorised personnel, including the protected service access steps. The spare TCB was flashed with a suitable firmware file, the configuration was not carried over in this case, and it took several reboots before the yellow warning fields cleared; firmware packages updated during the exchange and table calibration proved unnecessary. In the end the original board passed testing and was returned to service.

Safety

Isolate the table drive power and allow the VFD capacitors to discharge before touching the drive units, and complete the manufacturer's electrical safety and table motion (crush/pinch) checks before returning the scanner to clinical use.

Lesson: Test the original control board thoroughly before condemning it, and have the firmware, the configuration plan and the calibration expectations from the service documentation ready before starting any board swap.

Covered in the course: CT Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.