Canon/Toshiba · Aplio MX

Canon/Toshiba Aplio MX: console reports no printer despite working test page

What happened

Reported in the case: a Sony D898 MD printer on an Aplio MX was visible from Windows and printed a test page, but the ultrasound shell showed "no printer" and would not print from the console.

What was checked

Cause

The engineers found that the printer had never been assigned as the output device in the shell's own settings, so the application ignored a device the operating system saw as fully working.

What was done

The printer was selected as the output device in the console's output settings, after which printing from the console button worked. The way to reach this setting is a procedure from the manufacturer's documentation, done by authorised personnel.

Safety

Before disconnecting or reconnecting the printer, power down the system and unplug it from the mains, and confirm it passes electrical safety checks before returning to clinical use.

Lesson: A printer that works at the operating-system level can still be invisible to the application, so check the application's own output assignment before suspecting hardware.

Covered in the course: Ultrasound Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.